Recall Z-1698-2009 ยท FQH Lavage, Jet

Zimmer Inc.

Zimmer Pulsavac Wound Debridement System high capacity intramedullary brush tip, sterile, Zimmer, Dover, Ohio; Catatog Number: 00515018300. Used in conjunction with the Pulsavac Plus Wound Debridement System for pulsatile lavage/debridement.

DateJul 23, 2009 (initiated Jun 8, 2009)
StatusTerminated
ReasonThe seal on the sterile barrier pouch may be inadequate, resulting in lack of assurance of sterility.
Root causePackaging process control
ActionZimmer Orthopedic Surgical Products issued an "Urgent: Medical Device Recall" notice dated June 8, 2009 informing users of the affected devices and actions to be taken including the return of product and acknowledgement of receipt of notice by returning the Recall Certification Form by fax to 1-330-364-0974.
Quantity15
DistributionWorldwide Distribution -- United States, Canada and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.