Recall Z-1702-2009 ยท HSB Rod, Fixation, Intramedullary And Accessories

Zimmer Inc.

Zimmer ITST Intertrochanteric/subtrochanteric fixation taper reamer - Asia, nonsterile; REF 2258-50-01. (Note: This product is not distributed in the United States).

DateAug 4, 2009 (initiated Jun 5, 2009)
StatusTerminated
ReasonReamers labeled as having a standard diameter may actually be the dimensions of an Asia reamer, and vice versa.
Root causeProcess control
ActionZimmer sent an Urgent Device Recall letter dated 6/5/09 to all consignees and instructed to locate and return the products.
Quantity120 of both products
DistributionWorldwide distribution: USA, China, Germany, Japan, Korea and Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032367ITST INTERTROCHANTERIC SUBTROCHANTERIC INTRAMEDULLARY FEMORAL NAIL
Zimmer, Inc., Aug 12, 2003