Recall Z-2223-2009 ยท JJQ Colorimeter, Photometer, Spectrophotometer For Clinical Use

BioTek Instruments, Inc.

BioTek Gen5CL Microplate Data Collection and Analysis Software. Intended for clinical lab use with ELx800 and ELx808 microplate readers, offers preprogrammed clinical diagnostic assays within a PC-based application.

DateSep 18, 2009 (initiated Mar 12, 2009)
StatusTerminated
ReasonSoftware design allowed false negatives to be read when wells are dark.
Root causeSoftware design
ActionBioTek Instruments, Inc. issued a "Product Recall Notification" dated March 13, 2009 information customers of the affected product. Follow-up letters with software installation instructions were sent March 25, 2009. For further information, contact BioTek Instruments, Inc. at 1-888-451-5171 extension 405.
Quantity13
DistributionWorldwide Distribution -- United States (MN, CA, IL, MA, MI, CA, GA, FL, AZ and TX) and Peru.