Recall Z-0006-2010 ยท DKZ Enzyme Immunoassay, Amphetamine

Biosite Inc

Multi-Drug iCup Drug Screen Test Cup. Part Number DUD-1107-012-08A. Subsequent Product Codes: DIO, DIS, DJC, DJG, DJR, JXM, JXN, LCM, LDJ, LFG

DateOct 6, 2009 (initiated Apr 27, 2009)
StatusTerminated
ReasonThe cup associated with these lots of product may have a pin-sized hole in the floor of the cup. As a result, there is a potential for affected lots of cups to leak urine after an extended period of time.
Root causeProcess control
ActionBiosite, Inc. issued a "Notice of Correction" letter dated May 1, 2009 to consignees describing the problem and instructed users of proper use in order to avoid any cross contamination of samples until all affected product is completely used. For further information, contact Biosite, Inc. using the following information: Biosite Incorporated, an Inverness Medical Company 9975 Summers Ridge Road San Diego, CA 92121 Phone: 1-877-441-7440, option 2 Fax: 1-858-695-7100 Email: productadvisory@invmed.com
Quantity113,575 cups distributed in total
DistributionNationwide Distribution