Recall Z-1931-2009 ยท DSQ Ventricular (Assist) Bypass

Abiomed, Inc.

iPulse Circulatory Support System Catalog number: 0036-0010 The iPulse System Console is a multi-purpose mechanical circulatory support system used to support patients suffering from acute cardiovascular failure. The iPulse console can be used to drive exclusively one of two classes of therapeutic devices: either pneumatically actuated Ventricular Assist Devices (VADs) for total circulatory supp

DateSep 16, 2009 (initiated Apr 23, 2009)
StatusTerminated
ReasoniPulse Console SC1035 displayed a "Low Pressure / Low Flow" alarm and stopped pumping during patient transport in battery operation.
Root causeDevice Design
ActionAn "Urgent Voluntary Medical Device Field Correction Notice" was issued April 8, 2009 to affected customers. The customers were notified that the iPulse Console SC1032 displayed a "Low Pressure/ Low Flow" alarm and stopped pumping during patient transport in battery operation. Instructions were provided to users in case they encounter this failure mode. ABIOMED Field Service Engineers will contact customers to schedule a time to visit site and correct affected consoles. Please contact ABIOMED's Field Service at 1.800.554.8666, if you have any questions.
Quantity79
DistributionDomestic: MA, MI, SC, NY, NJ, OH, CA, PA, AR, MI, MD, CT, IL, WI, TX, NJ, TN, IN, MI, DE, FL, IA, WA, KY, and WV Foreign: Germany, Greece, Italy, Latvia, Spain, and Turkey
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.