ClearanceBrief
Recall Z-1828-2011 ยท RCE Cabinet X-Ray, Industrial

GE Inspection Technologies, LP

Phoenix x/ray nanotom series. Phoenix x|ray nanotom series cabinet x-ray systems are used in non-destructive testing, failure analysis, process control, and metrology.

DateApr 6, 2011 (initiated Mar 16, 2009)
StatusTerminated
ReasonSome of the systems do not have a safety interlock on the service access panel. Although the systems have two independent means for indicating when x-rays are being generated, some systems may have one indicator that is not discernible. Also, some systems lack proper placement of a caution label and others do not have this label affixed.
Root causeRadiation Control for Health and Safety Act
ActionThe firm has completed all of the following necessary actions and no further actions have been indicated: A Field Service Memo was issued to customers. It described the affected product and described the issue. GE Sensing and Inspection provided and installed retrofit kits, and the systems in question were tested by qualified Field Service Engineers at no charge to customers. GE asked customers to post the notification letter on the affected equipment until retrofit was completed. Questions were to be directed to the GE Service Center at 800-766-6062.
Quantity118 units total
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.