Recall Z-1794-2009 ยท FRN Pump, Infusion

Cardinal Health 303 dba Cardinal Health

Alaris System Inter-Unit Interface (IUI) Connectors. The Modules include the following: Alaris PC Unit, Model Numbers 8000 and 8015; Alaris Pump Module, Model 8100; Alaris Patient-Controlled Analgesia (PCA), Model 8120; Alaris Syringe Module, Model 8110; Alaris SpO2 Module, Model 8210 and 8220 and Alaris EtCO2 Module, Model 8300.

DateAug 3, 2009 (initiated Jun 12, 2009)
StatusTerminated
ReasonThere is a potential risk of IUI connector failure in the Alaris System when these connectors become corroded or contaminated, which may result in an interruption of infusion with an audible and visual alarm. IUI connectors can become corroded or contaminated when exposed to fluid spills or improper cleaning. This notification is to reinforce the importance of inspection of IUI connectors prior
Root causeDevice Design
ActionThe recall was initiated on June 12, 2009 with the firm forwarding via UPS an Urgent: Medical Device Recall Notification with attached Customer Response Card to the Chief Administrative Officer, Director of Pharmacy, Director of Biomedical Engineering, Director of Nursing & Director of Risk Management of all of the affected consignees. Customers were also asked to please promptly complete and return the enclosed Customer Response Card to expedite the correction process. The Recall Notification informed the consignees of: (A) PROBLEM #1: Occlusion Warning Message for Alaris Pump Module; (B) PR
Quantity500,000 units
DistributionWorldwide Distribution -- US (including Puerto Rico and Guam), Germany, England, Canada, Netherlands, Malaysia, Singapore, Australia, and New Zealand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012383GEMINI PC-1,PC-2, PC-2TX,PC-4, SIGNATURE EDITION INFUSION PUMP (SE), MEDSYSTEM III INFUSION (MSIII), MEDSYSTEM PATIENT C
Alaris Medical Systems, Inc., Oct 2, 2001
K023264MEDLEY SYRINGE PUMP MODULE SYSTEM, MODEL 8110
Alaris Medical Systems, Inc., Dec 19, 2002
K032233MEDLEY PCA MODULE, MODEL 8120
Alaris Medical Systems, Inc., Sep 9, 2003
K051641ALARIS SYSTEM PC UNIT, MODEL 8001
Cardinal Health, Alaris Products, Oct 20, 2005
K950419THE IMED ORION INFUSION PUMP AND ADMINISTRATION SETS
Imed Corp., Jun 21, 1995