ClearanceBrief
Recall Z-1793-2009 ยท FRN Pump, Infusion

Cardinal Health 303 dba Cardinal Health

Alaris Pump Module, Model 8100.

DateAug 3, 2009 (initiated Jun 12, 2009)
StatusTerminated
ReasonThere is a potential risk of over infusion when certain fluids solidify in the pumping mechanism of the Alaris Pump module. This notification is to reinforce the importance of proper techniques for IV set loading and cleaning. When IV sets are improperly loaded, the set can tear and cause medication to leak and potentially enter the bezel area. Improper cleaning of the device, such as using hig
Root causeDevice Design
ActionThe recall was initiated on June 12, 2009 with the firm forwarding via UPS an Urgent: Medical Device Recall Notification with attached Customer Response Card to the Chief Administrative Officer, Director of Pharmacy, Director of Biomedical Engineering, Director of Nursing & Director of Risk Management of all of the affected consignees. Customers were also asked to please promptly complete and return the enclosed Customer Response Card to expedite the correction process. The Recall Notification informed the consignees of: (A) PROBLEM #1: Occlusion Warning Message for Alaris Pump Module; (B) PR
Quantity300,000 units
DistributionWorldwide Distribution -- US (including Puerto Rico and Guam), Germany, England, Canada, Netherlands, Malaysia, Singapore, Australia, and New Zealand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K950419THE IMED ORION INFUSION PUMP AND ADMINISTRATION SETS
Imed Corp., Jun 21, 1995