Recall Z-1545-2009 ยท IYE Accelerator, Linear, Medical

Brainlab AG

BrainLab m3 Micro-Multileaf-Collimator for Varian Type III with MLC, model 41600-3; BrainLab AG, Kapellenstrasse 12, 85622 Feldkirchen, Germany. The mMLC is intended to assist the radiation oncologist in the delivery of radiation to well-defined target volumes while sparing surrounding normal tissue and critical organs from access radiation.

DateJul 2, 2009 (initiated Mar 19, 2009)
StatusTerminated
ReasonThe Integrated Multileaf-Collimator (MLC) of the linear accelerator could interfere with the intended conformal fields of the BrainLab m3 micro-Multileaf-Collimator (mMLC) during patient treatment.
Root causeOther
ActionBrainLAB AG issued an "Urgent Field Safety Notice/Product Notification" letter dated March 19, 2009 informing all customers of the affected device. Consignees were instructed on proper steps to take to avoid further issues. In addition, the most current m3 mMLC user guide, version 7.4, was scheduled to be sent to users in April 2009. For further questions, contact BrainLAB AG by phone at 1-800-597-5911).
Quantity35 units
DistributionNationwide Distribution - AL, CA, CO, FL, GA, IL, KS, KY, LA, MD, MI, MN, MS, ME, NJ, NY, OR, TN, TX, UT, VA, WV, WA and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K004022M3 (MIRCO-MULTILEAF COLLIMATOR