ClearanceBrief
Recall Z-1790-2009 ยท FRN Pump, Infusion

Cardinal Health 303 dba Cardinal Health

The Alaris PC Unit, Model Numbers 8000 and 8015 (formerly Medley PC Unit) with software version 4 and above.

DateAug 3, 2009 (initiated Jun 12, 2009)
StatusTerminated
ReasonThere is a potential risk for medication not being delivered when IV tubing sets are occluded in the Alaris Pump module. This can occur if IV sets are set up in the infusion pumps but not used for an extended period of time. If users fail to clear the occlusion, then the occluded tubing may cause medication not to be delivered. A delay of medication may result in serious injury and/or death, an
Root causeDevice Design
ActionThe recall was initiated on June 12, 2009 with the firm forwarding via UPS an Urgent: Medical Device Recall Notification with attached Customer Response Card to the Chief Administrative Officer, Director of Pharmacy, Director of Biomedical Engineering, Director of Nursing & Director of Risk Management of all of the affected consignees. Customers were also asked to please promptly complete and return the enclosed Customer Response Card to expedite the correction process. The Recall Notification informed the consignees of: (A) PROBLEM #1: Occlusion Warning Message for Alaris Pump Module; (B) PR
Quantity186,000 units
DistributionWorldwide Distribution -- US (including Puerto Rico and Guam), Germany, England, Canada, Netherlands, Malaysia, Singapore, Australia, and New Zealand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012383GEMINI PC-1,PC-2, PC-2TX,PC-4, SIGNATURE EDITION INFUSION PUMP (SE), MEDSYSTEM III INFUSION (MSIII), MEDSYSTEM PATIENT C
Alaris Medical Systems, Inc., Oct 2, 2001
K051641ALARIS SYSTEM PC UNIT, MODEL 8001
Cardinal Health, Alaris Products, Oct 20, 2005