Recall Z-1516-2009 ยท DXY Implantable Pacemaker Pulse-Generator
Medtronic Inc. Cardiac Rhythm Disease Managment
Medtronic SIGMA 100/200/300 Series IPGs, Dual Chamber Rate Responsive Pacemaker, models DR353, SDR203, SDR303, and SDR306.
| Date | Jun 11, 2009 (initiated May 18, 2009) |
| Status | Terminated |
| Reason | One or more bond wire pairs will lift or separate from the bonding terminals on the device hybrid. This may present clinically as loss of rate response, premature battery depletion, loss of telemetry, or no output. |
| Root cause | Device Design |
| Action | Medtronic Sales representative hand delivered an Important Patient Safety Information letter, dated May 2009, to physicians beginning 05/18/09. The letter describes the issues, provides a predicted failure rate for the 3 populations of devices, and provides Patient Management Recommendations. The letter recommended that physicians should advise their patients to seek medical attention immediately if they experience symptoms (e.g., fainting or lightheadedness). It was also recommended that physicians should consider device replacement for patients who are both pacemaker dependent and who ha |
| Quantity | 12891 |
| Distribution | Worldwide Distribution -- USA, including Washington D.C., Guam, Puerto Rico, and Virgin Islands; Albania, Algeria, Argentina, Armenia, Aruba, Australia, Austria, Bahamas, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyp |