Recall Z-1921-2009 ยท KWP Appliance, Fixation, Spinal Interlaminal

SeaSpine, Inc

Malibu Polyaxial Screw, 6.5 x 45mm Part Number 12-6545 Malibu Spinal System when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System is to provide immobilization in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine.

DateSep 3, 2009 (initiated Mar 4, 2009)
StatusTerminated
ReasonSeaSpine has received reports relating to rust on the internal washer contained within the titanium seat assembly of certain Malibu Polyaxial and Malibu Reduction Screws.
Root causeMixed-up of materials/components
ActionIn an Urgent Voluntary Product Recall Notification letter dated March 5th, 2009, SeaSpine informed consignees about the reason for recall and asked them to immediately check the products in their possession and hospital inventories to determine if they have any of the affected product. Alternatively, there were told if they would prefer to return to SeaSpine all Malibu Screw inventory that is in their possession, SeaSpine will sort out the affected product upon receipt and will replace their inventory with Malibu Screws that conform to all SeaSpine specifications. The part number and lot are
Quantity1,116 Screws in total
DistributionWorldwide Distribution -- USA, including states of AL, AZ, CA, CO, FL, GA, IN, MD, MI, MO, MS, NM, OK, TN, TX, VA, and WI and country of Greece.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051942MALIBU SPINAL SYSTEM
Seaspine, Inc., Jul 27, 2005