Recall Z-1569-2009 ยท BRY Cabinet, Table And Tray, Anesthesia

Stryker Communications Corp

Amico E-Cylinder Canister Holder, distributed by Stryker Communications, Flower Mound, TX 75028. Canister holder is used to hang E-Cylinders (pressurized gas canisters) from a boom. Product typically used in the surgical setting.

DateJul 13, 2009 (initiated May 19, 2009)
StatusTerminated
ReasonDesign flaw on the E-Cylinder canister holder may result in early fatigue and failure of the bracket screws, which could cause the cage to become detached from its mount and fall.
Root causeDevice Design
ActionAn "Urgent: Device Recall" letter dated May 19, 2009 was issued to customers. Customers were advised to discontinue use of all affected E-cylinder holders. Consignees were asked to return the "Product Accountability Form" included with the recall letter, after which consignees will be provided with an improved cylinder holder at no cost. Direct questions to Stryker Communications at 1-972-410-7100.
Quantity66 units.
DistributionNationwide and Canada.