Recall Z-1677-2009 ยท KPS System, Tomography, Computed, Emission

Siemens Medical Solutions USA, Inc.

Symbia T Series SPECT-CT (Single-Photon Emission Computed Tomography and Computed Tomography) System; Siemens Medical Solutions USA, Inc., Hoffman Estates, IL 60195-2061 Indications for use: SPECT + CT: Perform scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's

DateAug 4, 2009 (initiated Apr 22, 2009)
StatusTerminated
ReasonThe pinion gear could crack and cause the gantry to spin freely thereby damaging the Symbia S or T System and potentially injuring a patient during a scan.
Root causeFinished device change control
ActionSiemens sent Urgent Field Correction Recall letters dated 4/22/09 to the affected users of the Symbia S and Symbia T Systems. The accounts were informed of the potential faulty pinion gear which could crack and cause the gantry to spin freely, damaging the system and potentially injuring a patient during a scan. The customers were advised to stop using the system if they hear any abnormal grinding noise from the gantry. They also were advised that a Siemens Customer Service Representative would contact them within 60 days to schedule a replacement of the pinion gear free of charge. Any ques
Quantity38 units
DistributionWorldwide Distribution -- USA, Australia, Austria, Brazil, Bulgaria, Canada, China, Finland, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Philippines, Russia, Spain, Taiwan, and Turkey.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041166SIEMENS ENHANCED IMAGING SYSTEM
Siemens Medical Solutions USA, Inc., May 17, 2004