Recall Z-1612-2009 ยท MNH Orthosis, Spondylolisthesis Spinal Fixation

Stryker Spine

Stryker Mantis Rod Inserter Inner Shaft; Non Sterile; Manufactured by Stryker Spine SAS, Cestas France. Distributed by Stryker Spine, Allendale, NJ 07401-1677. The MANTIS Spinal System is intended for posterior, non-cervical pedicle fixation for the following indications: degenerative disk disease, spodylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudoarthrosis and failed previous

DateJul 7, 2009 (initiated Apr 30, 2009)
StatusTerminated
ReasonThe Mantis Rod Inserter Shaft was found to have deformation of the stainless stell split sphere when subjected to high forces, prohibiting the user from being able to grip the rod implantation and the split sphere may separate and fall into the patient.
Root causeDevice Design
ActionStryker Spine issued an "Urgent Product Recall" notification dated April 29, 2009 via Federal Express to all Consignees. Users were instructed to quarantine all affected product and fax a copy of the Product Accountability Form to 201-760-8406. A Stryker sales representative will retrieve the quarantined product. For further information, contact Stryker Spine at 201-760-8206.
Quantity387 in US
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061813MANTIS SPINAL SYSTEM
Stryker Spine, Sep 19, 2006