Recall Z-1657-2009 ยท MGB Device, Hemostasis, Vascular

Abbott Vascular-Cardiac Therapies dba Guidant Corporation

Abbott Vascular, StarClose Vascular Closure System Part Number 14677-01, for closure of femoral artery access sites after cardiovascular catheterization.

DateJul 6, 2009 (initiated May 6, 2009)
StatusTerminated
ReasonPotential sterility issue. Internal component quarantined for possible contamination, was released for sale after going through sterilization procedure.
Root causeProcess design
ActionBetween May 6 and May 8, 2009, Sales Representatives visited accounts to recover the product. Recall notifications (Urgent Product Recall letters, dated May 6, 2009) were hand delivered to customers advising them Abbott Vascular is voluntarily recalling one lot of the StarClose Vascular Closure System for a potential stenlity issue discovered internally. Customers are asked to cease use of devices from this lot and work with their local account representative to review their inventory. complete the attached Recall Effectiveness Check Form and rerturn Identified products to Abbott Vascular.
Quantity240 units
DistributionNationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.