ClearanceBrief
Recall Z-0137-2010 ยท IYO System, Imaging, Pulsed Echo, Ultrasonic

GE Medical Systems, LLC

GE Healthcare, GE Innova 4100 / 4100 IQ. , ( The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector)

DateNov 9, 2009 (initiated Apr 16, 2009)
StatusTerminated
ReasonPotential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room monitor stating that three minutes are left before the X-ray is inhibited. This could lead to a delay in treatment and possible additional X-ray exposure and contra
Root causeDevice Design
ActionConsignees were sent on May 1, 2009 a GE Healthcare " Urgent Medical Device Correction" letter dated April 16, 2009. The letter was addressed to Hospital Administrator/ Risk Manger, Mangers of Radiology/ Cardiology, and Radiologists/ Cardiologists. The letter described the problem, Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information.
Quantity35
DistributionWorldwide distribution: USA, ARGENTINA, BELGIUM, BRAZIL, BULGARIA, CHINA, CROATIA, EGYPT, FRANCE, GERMANY, HONG KONG, HUNGARY, INDIA, ITALY, JAPAN, REPUBLIC OF KOREA, LATVIA, LIBYA, MEXICO, NORWAY, POLAND, RUSSIA, SINGAPORE, SPAIN, SWEDEN, TAIWAN, TURKEY, UNITED KINGDOM, ROMANIA, ISRAEL, UKRAINE
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023178INNOVA 4100
Ge Medical Systems, Inc., Nov 26, 2002
K033244DIGITAL FLUOROSCOPIC IMAGING SYSTEM, MODEL INNOVA 4100 WITH TILT TABLE OPTION
Ge Medical Systems, Inc., Oct 16, 2003
K042053INNOVA 4100, INNOVA 3100
Ge Medical Systems, Inc., Aug 26, 2004
K052157INNOVA CT OPTION FOR DIGITAL FLUOROSCOPIC IMAGING SYSTEMS INNOVA MODELS 4100, 4100IQ, 3100, 3100IQ & 2100IQ
Ge Medical Systems, LLC, Aug 25, 2005
K052412INNOVA 2100-IQ, INNOVA 3100, INNOVA 3100-IQ, INNOVA 4100, AND INNOVA 4100-IQ
GE Healthcare, Sep 16, 2005
K061163INNOVA 2121IQ, INNOVA 3131IQ, INNOVA 4100, INNOVA 4100IQ, INNOVA 3100, INNOVA 3100IQ, INNOVA 2100IQ WITH INNOVA IVUS OPT
Ge Medical Systems, LLC, Jun 7, 2006