Recall Z-1906-2009 ยท JQP Calculator/Data Processing Module, For Clinical Use

Perkin Elmer

Specimen Gate Lab - MSMS Data Suite, Product Code: 5002-0310. The intended use of the product is data management in prenatal and neonatal screening laboratories designed to allow viewing and management of assay results.

DateAug 25, 2009 (initiated Jan 15, 2009)
StatusTerminated
ReasonIf Result Codes are used and if they are manually edited, they are correct unless the laboratory specifically uses the View Assay button in the Specimen History screen to open another assay, and then returns back to the original assay
Root causeSoftware design
ActionPerkin Elmer began notifying US customers January 15, 2009. Notification included instructions for how to avoid the situation and instructions for how to obtain a correction. The firm requested consignees to send in a Response Form by fax (358-2-2678-357) or by email (TurkuQMresponse@perkinelmer.com). For further information, please contact your closest Specimen Gate support center at specimen.gate.support@perkinelmer.com for North America and specimengatesupportfi@perkinelmer.com for outside of North America.
Quantity13 units
DistributionWorldwide Distribution -- United States (AL, AZ, CA, CO, CT, FL, GA, IN, KS, KY, MI, MO, OH, OK, OR, TN and TX) and Canada.