Recall Z-0469-2011 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips Medical Systems North America Co. Phillips

The product's marketing brochure states: "1.5T coils and accessories...Philips Achieva and Intera 1.5T MR systems". The Identification of the Synergy Flex-M/ Shoulder Coil 1.5T can be found on the driver box. The 12nc of the affected system is labeled as 4522-131-6656x.

DateNov 26, 2010 (initiated May 1, 2009)
StatusTerminated
ReasonCombined use of the Synergy Flex-M / Shoulder Coil 1.5T increases the chance of RF interaction and heating up of the coil. This may result in possible burns of the patient.
Root causeOther
ActionPhilips issued URGENT-Field Safety Notice letters dated May 1, 2009 and October 14, 2009 to their consignees. The letter informed customers that the firm will replace the "old" Flex M/Shoulder Coils with a new Flex-M coil. Consignees could contact Philips Healthcare Call Center at 800-722-9377, #5, #3, #1 and reference FCO 78100290 for any questions.
Quantity389 units
DistributionNationwide Distribution: in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VA, WA, WI, WV, WY, and Washington DC. There were no foreign consignees.