Recall Z-1389-2012 ยท NEP System/Device, Pharmacy Compounding

Baxa Corporation

Pharmacy compounding system . It is a twenty-four source compounder for multi-source fluid mixing of both macro and micro ingredient down to 0.2 ml quantities. Product Usage: Pharmacy compounding System for parenteral macro and micro ingredients

DateApr 9, 2012 (initiated Mar 23, 2009)
StatusTerminated
ReasonSoftware issue with compounding pump resulting in over delivery of component ingredients during routine operations.
Root causeSoftware design
ActionBaxa sent a Safety Alert letter dated March 23, 2009 to all affected consignees. The letter identified the affected product, problem and recommended actions with specific instructions related to the issue. Additionally, the firm will be providing a warning label for the affected product as soon as possible and is currently working on a product upgrade to correct the issue. For assistance contact Baxa Technical Support at 800-678-2292. The firm is also providing a 24 hour a day, 7 days a week Technical Support for the affected product.
Quantity1322 units
DistributionWorldwide Distribution - USA Nationwide including Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.