Recall Z-1558-2009 ยท MSX System, Network And Communication, Physiological Monitors
GE Medical Systems, LLC
Datex-Ohmeda S/5 Device Interfacing Solution, N-DISVENT 02. The Datex-Ohmeda S/5 Device Interfacing Solution, N-DISxxxx 01 with N-DISVENT 02, is indicated for data transfer between stand-alone monitors, ventilators/anesthesia machines, blood gas analyzers, and heart-lung machines and Datex-Ohmeda bedside monitors for displaying and patient care information purposes. The device is intended for us
| Date | Sep 4, 2009 (initiated May 8, 2009) |
| Status | Terminated |
| Reason | GE Healthcare has recently become aware of an issue in the calculations associated with Datex-Ohmeda S/5 Device Interfacing Solution N-DISVENT-02 interfaced between certain Datex-Ohmeda patient monitors (Datex-Ohmeda S/5 Anesthesia Monitor (AM), Datex-Ohmeda S/5 Critical Care Monitor (CCM), Datex-Ohmeda S/5 Compact Anesthesia Monitor (CAM), Datex-Ohmeda S/5 Compact Critical Care Monitor (CCCM)) an |
| Root cause | Software design |
| Action | An "Urgent Medical Device Correction" letter was issued to consignees on or about April 9, 2009. The letter was addressed to Healthcare Administrator/Risk Manager, Chief of Nursing, Director of Biomedical Engineering. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information. A representative from GE Healthcare will contact customers to schedule implementation of the correction when it is available for installation at the customer's facility. Direct questions regarding this notice to GE Healthcare Technical Support or |
| Quantity | 2,638 |
| Distribution | Worldwide Distribution -- US (including DC and states of AL, AK, AZ, AR, CA, CO, FL, IL, IN, IA, KS, KY, LA, ME, MD, MI,MN, MO,MT, NE, NH, NJ, NY, NC, OH, OK OR, PA, SC, TX, UT,VT, VA, WA, WV and WI) and countries of AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, CANADA, CHINA, CYPRUS, DENMARK, ESTONIA, FRAN |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K070421 | DATEX-OHMEDA S/5 DEVICE INTERFACING SOLUTION, N-DISXXXX..01 WITH N-DISVENT..02 GE Healthcare, Mar 13, 2007 |