ClearanceBrief
Recall Z-1556-2009 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Medical Systems, LLC

GE Signa Ovation 0.35 T MR System. The Signa Ovation System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times.

DateJun 30, 2009 (initiated Apr 16, 2009)
StatusTerminated
ReasonGE Healthcare has identified a potential safety issue with the Signa OpenSpeed and Signa Ovation MR Scanner tables that may impact patient safety. The table may begin lowering on its own while a patient is on the table. If this occurs while the table is in the magnet bore, the table may tilt and could result in patient injury. The patient's head or feet depending on patient orientation, may hit
Root causeDevice Design
ActionGE Healthcare issued an "Urgent Medical Device Correction" letter dated April 16, 2009 addressed to Hospital Administrators, Mangers of Radiology/Cardiology and Radiologists/Cardiologists. The letter describes the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information. A GE Healthcare Field Engineer will schedule a field visit to correct the device. Further questions or concerns may be addressed to GE Healthcare at 1-262-521-6681.
Quantity203
DistributionWorldwide Distribution -- US including PR and states of AL, AZ, AR, CA, CO,CT, DE, FL, GA, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MN, MS, MO,MT, NV, NH, NE, NJ, NY, NC, OH, OK,OR, PA, SC, SD, TN, TX, VA, WA, WV, WI and countries of BRAZIL CHINA, DENMARK, EGYPT, ENGLAND, GERMANY, GUATEMALA, HU
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K053009GE 0.35T SIGNA OVATION WITH EXCITE MR SYSTEM
GE Medical Systems, Nov 17, 2005