Recall Z-0528-2010 ยท KPR System, X-Ray, Stationary

GE Medical Systems, LLC

GE Healthcare Definium 8000.Model # 5131070. GE Healthcare, 3000 North Grandview, Waukesha, WI 53188. The Revolution XR/d with Tomosynthesis is intended for use in generating Tomographic Images of human anatomy.The device is not intended for mammographic applications.

DateDec 16, 2009 (initiated Aug 13, 2009)
StatusTerminated
ReasonCertain GE Radiographic fixed X-ray tables (Proteus XR/a, Revolution XR/d, and Definium 8000) were found to have tabletops that move unexpectedly in a bi-directional mode. In rare cases, a malfunction may occur, which could result in the tabletop brakes falling to engage when the control foot pedal is released to lock the floating tabletop in place.
Root causeNonconforming Material/Component
ActionA GE Healthcare "Urgent Medical Device Correction" letter dated April 9, 2009 was issued to customers. The letter was addressed to Hospital Administrators/Managers of Radiology and Radiologists. The letter described the Product Issue, Affected Product, Safety Instructions, Product Correction and Contact Information. Operators should verify that the tabletop is locked before the patient is loaded on or off the table. A Field Modification will be performed on the affected products. A GE Healthcare Field Engineer will schedule a field visit to make this modification. Direct questions or c
Quantity5,943 of total product
DistributionWorldwide Distribution -- UNITED STATES (including Puerto Rico), VIETNAM, VENEZUELA, URUGUAY, UNITED KINGDOM, UNITED ARAB EMIRATES, UKRAINE, TURKS AND CAICOS ISLANDS, TURKEY, TUNISIA, TRINIDAD AND TOBAGO, THAILAND, TAIWAN, SWITZERLAND, SWEDEN SPAIN, SOUTH AFRICA, SLOVENIA, SLOVAKIA, SINGAPORE, SAUDI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051967REVOLUTION XR/D DIGITAL RADIOGRAPHIC IMAGING SYSTEM WITH TOMOSYNTHESIS
GE Healthcare, Aug 9, 2005