Recall Z-1554-2009 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Radiometer America Inc

Waste bottles Part Numbers/Article Numbers: 905-590/D512 and 905-802/D513. The waste bottles are used to collect waste from the ABL7XX and the ABL8XX blood gas analyzer after rinsing.

DateJul 15, 2009 (initiated Apr 16, 2009)
StatusTerminated
ReasonWhen removing the waste bottle from the analyzer waste material may leak both from the analyzer and from the waste bottle.
Root causeProcess design
ActionRadiometer America Inc. issued a recall notification letter dated May 5, 2009 to all customers containing information on replacement of all potentially defective waste containers and instruction on how to handle any mounted waste containers. For further questions, contact Radiometer America Inc. at 1-800-736-0600.
Quantity905-590 (D512): 35,552 bottles; 905-802 (D513): 45,204 bottles
DistributionWorldwide Distribution to US and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002290ABL-700 SERIES, MODEL ABL 7XX (VARIOUS MODELS)
Radiometer Medical A/S, Sep 13, 2000
K041874ABL800 FLEX
Radiometer Medical Aps, Oct 8, 2004