Recall Z-1436-2009 ยท DXJ Display, Cathode-Ray Tube, Medical

Philips Healthcare Inc.

CareVue Chart Release C.O Part Number: 862246 Intended for use in data collection, storage, and management with independent bedside devices, and ancillary systems that are connected either directly or through networks. It is indicated for use by healthcare providers whenever there is a need for generation of a patient record and computation of drug dosage.

DateMay 27, 2009 (initiated May 5, 2009)
StatusTerminated
ReasonSoftware: A software defect under certain circumstances could lead to the unintended administration of a duplicate dose of a scheduled medication within a single 24 hour period
Root causeSoftware design
ActionPhilips Medical notified accounts by letter on 5/6/09 the Urgent - Medical Device Correction notification. While customers await the correction of their devices, customers are given the following instructions: Clinicians should not rely solely on the worklist within CareVue Chart to determine actionable orders for the patient until their software is upgraded. Clinical users should always use the medication administration record (MAR) and the administration record in determining the care their patients receive in addition to the critical care worklist. Philips is currently developing a correct
Quantity103 units
DistributionWorldwide Distribution -- USA including states of CA, FL, MA, OH, OK, RI, VT, and WV and countries of Canada, Australia, Austria, Belgium, Denmark, Finland, France, Germany, Italy, Malaysia, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, and the United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K992636HP CAREVUE 9000 CLINICAL INFORMATION SYSTEM, MODEL M2331A
Hewlett-Packard Co., Aug 31, 1999