Recall Z-1654-2009 ยท DXJ Display, Cathode-Ray Tube, Medical

Philips Healthcare Inc.

Philips IntelliVue Clinical Information Portfolio Software Version D.0, for patient data collection, storage, and management.

DateJul 7, 2009 (initiated May 5, 2009)
StatusTerminated
ReasonError in therapy scheduling: If scheduled orders are documented and edited prior to the orders' original schedule time, ICIP can create a second instance of the same order marked as pending. A second instance of an order may result in a patient receiving duplicate therapy or patients may not receive their scheduled therapies.
Root causeSoftware design
ActionPhilips notified customers by Urgent Medical Device Correction Notification letter on May 6, 2009. The Customer notifications in the United States will be delivered via UPS. Customer notifications outside the US will be managed by Philips representatives in each affected geography. Customers will be asked to follow the guidelines in the Action to be taken by Customer/User section of the Urgent Medical Device Correction Notice until they receive their software corrections. The corrections will be in the form of software upgrades and will be performed by Philips service personnel. On May 27, 20
Quantity105 units
DistributionWorldwide distribution, including USA, Australia, Belgium, China, Denmark, France, Germany, Ireland, Italy, Kenya, Malaysia, Netherlands, Singapore, Spain, Switzerland, Taiwan, UAE, and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K992636HP CAREVUE 9000 CLINICAL INFORMATION SYSTEM, MODEL M2331A
Hewlett-Packard Co., Aug 31, 1999