Recall Z-1644-2009 ยท NGY Cannula, Trocar, Ophthalmic

Alcon Laboratories, Inc

Alcon Vitrectomy 25 GA Custom Pak, Catalog/Item #10240-03, 11077-03, 11187-02 and 13305-01; distributed by Alcon Laboratories, Inc., Fort Worth, TX 76134.

DateMay 26, 2009 (initiated Apr 29, 2009)
StatusTerminated
ReasonSpecific lots of Alcon 25 GA Trocar Blades do not conform to finished product specifications in that oxidation can occur on some of the blades.
Root causeProcess control
ActionFirm began notifying consignees of recall by Urgent: Medical Device Recall letter on 04/29/09 (letters sent via FedEx). Consignees instructed to check inventory for affected lots; to complete and return the Medical Device Recall Response Form; to contact Alcon to arrange for return of the devices; that credit will be issued for the recalled product when it is returned; and in the case that there is no inventory remaining, to indicate that on the Response Form and return it as well. Consignees should contact Alcon Technical Consumer Affairs at 1-800-561-6466, Option 8 for technical assistance
Quantity59 surgical paks.
DistributionWorldwide Distribution -- USA and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K880961ALCON SURGICAL PROCEDURE PACKS