Recall Z-1694-2009 ยท MNH Orthosis, Spondylolisthesis Spinal Fixation
X Spine Systems Inc
Capless Li Pedicle Screw System, Diameters - 4.75mm, 5.5mm, 6.5mm, 7.5mm, 8.25mm, Lengths 30mm-55mm, labeled non-sterile. The Capless LI Pedicle Screw System is indicated for the treatment of sever spondylolisthesis of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft.
| Date | Aug 19, 2009 (initiated Mar 24, 2009) |
| Status | Terminated |
| Reason | A defect cause was discovered following a customer complaint pertaining to a post operative dissociation of the Capless Li Screw construct. The screw assembly is made up of three parts; a cup which locks onto a 5.5mm rod, a yoke that cradles the rod during fusion, and a screw that attached to the vertebra. The screw ball on the screw was manufactured out of tolerance (undersized). |
| Root cause | Process control |
| Action | X Spine Systems Inc. issued advisory notices in March 2009 to Consignees notifying them of the problem and requesting all affected product be returned to the firm. The firm also conducted telephone calls and emails for the return of the product. For further information, contact X Spine Systems Inc. at 1-937-847-8400 extension 115. |
| Quantity | 9,613 units |
| Distribution | Worldwide Distribution -- Including states of AZ, CA, CO, FL, ID, IN, MO, MS, NV, OK, TN, TX, UT and WY and countries of Spain, South Africa and the United Kingdom. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K072282 | CAPLESS LI PEDICLE SCREW SYSTEM X-Spine Systems, Inc., Sep 14, 2007 |