ClearanceBrief
Recall Z-1439-2009 ยท IZI System, X-Ray, Angiographic

Siemens Medical Solutions USA, Inc

Mavig Portegra 2 which is an accessory to the Siemens angiographic, fluoroscopic and radiographic imaging systems. Component material number 7559375 Upper Body radiation shield, Artis T.

DateJun 16, 2009 (initiated May 1, 2009)
StatusTerminated
ReasonImproper installation of accessory support arm, which could potentially result in the loosening of the arm over time causing it to fall.
Root causeOther
ActionThe recalling firm has decided to inspect all of the installed units for this condition and where necessary replace the missing clip. The recalling firm has issued an Update Instructions AX011/09/S dated 03.09, which provides instructions on how to perform a visual inspection of the Mavig Portegra 2 arm for the presence of the safety clip and the installation instructions on how to properly install the clip if it is missing.
Quantity427 units
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010721AXIOM ARTIS MODULAR ANGIOGRAPHY SYSTEM
Siemens Medical Corp., Mar 30, 2001