ClearanceBrief
Recall Z-1536-2009 ยท LXT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component, Metal Composite

Zimmer Inc.

TransFx External Fixation System Drill Bit, Quick Connect, 4.0 mm diameter, 195 mm length, nonsterile, Zimmer, Warsaw, IN; REF 4450-63. TransFx drill bits can be found in the following kits: TransFx Intermediate External Fixation System, REF 4450-05-10; TransFx Large External Fixation System, REF 4450-010-10 and TransFx Small External Fixation System, REF 4450-00050-15. For use in pre-drilling

DateJun 30, 2009 (initiated May 1, 2009)
StatusTerminated
ReasonDrill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation.
Root causeProcess control
ActionDistributors and consignees were notified by an Urgent: Device Recall letter dated 5/1/09. Consignees were advised to stop using the device and quarantine it immediately. Distributors were instructed to carry out a physical count of all affected product and record the data on the Inventory Return Certification Form; fax a copy of the completed form to Zimmer, Inc.; and return recalled product along with the Inventory Return Certification Form to the firm.
Quantity2146
DistributionWorldwide Distribution -- USA, Australia, Canada, El Salvador, England, Germany, Latin America, and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003205TRANSFX ANGLED RODS
Immedica, Inc., Dec 15, 2000
K984357TRANSFX EXTERNAL FIXATION SYSTEM
Immedica, Inc., Mar 4, 1999
K990848TRANSFX INTERMEDIATE EXTERNAL FIXATION SYSTEM,TRANSFX SMALL EXTERNAL FIXATION SYSTEM, TRANSFX MINI EXTERNAL FIXATION
Immedica, Inc., May 17, 1999
K991723TRANSFX MULTI PIN CLAMP
Immedica, Inc., Jul 22, 1999