Recall Z-1391-2009 ยท MMI Immunoassay Method, Troponin Subunit

Biosite Inc

Triage Cardiac Panel, Biosite Catalog Number 97000HS, 25 individually pouched devices in a labeled kit box. Fluorescence immunoassay kit to be used with the Triage Meter for the quantitative determination of myoglobin, creatine kinase MB fraction (CK-MB), and troponin I in EDTA whole blood and plasma specimens. These three cardiac biomarkers are used as an aid in the diagnosis of myocardial inf

DateMay 8, 2009 (initiated Apr 27, 2009)
StatusTerminated
ReasonThe recall was initiated because the use of the affected lot may lead to reporting falsely low values for troponin I and myoglobin, possibly resulting in missed or incorrect diagnosis.
Root causeProcess control
ActionThe recall was initiated with "Urgent Medical Device Recall" notices, dated 4/28/2009, with attached "Customer/Distributor Verification Form" being sent to the affected customers via fax, e-mail or direct mail on 04/28/2009. Customers were informed of the reason for the recall and instructed to immediately discontinue all use of the listed products and discard all affected product in accordance with your their regulations. The customers were give two options; "Option 1: Complete and FAX the enclosed Verification Form within 10 days to confirm your receipt of this notice, to indicate the num
Quantity401 kits
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030286TRIAGE CARDIO PROFILER/TRIAGE CARDIAC PANEL
Biosite Incorporated, Feb 21, 2003