Recall Z-1656-2009 ยท MKJ Automated External Defibrillators (Non-Wearable)

Welch Allyn Protocol, Inc

Welch Allyn AED 10 automatic external defibrillator and MRL Jumpstart (collectively "AED 10")

DateJul 6, 2009 (initiated Apr 24, 2009)
StatusTerminated
ReasonPossible for users to misunderstand signals. The original product labeling told users that a flashing low battery indicator signals that the battery is low and requires changing soon. If a user experiences a flashing low battery status indicator and does not understand that the device can continue to be used, the user might choose not to continue operating the device, thereby causing therapy to
Root causeLabeling design
ActionThe firm has developed expanded labeling for users and will revise the user manuals for software versins 2.03, 2.04, 2.05, and 2.06. Customers receiving the updated manuals will be sent a cover letter directing them to discard the original user manual and replace it with the updated manual. The April 24, 2009, cover letter will highlight the clarification of directions for use regarding what to do when a low battery indicator is displayed.
Quantity46740 units
DistributionWorldwide distribution: USA, Guam, Puerto Rico, Canada, Mexico, Andorra, UAE, Afghanistan, Australia, Azerbaijan, Belgium, Belarus, Bermuda, Bahrain, Brazil, British West Indies, Switzerland, China, Cyprus, Czech Republic, Germany, Denmark, Algeria, Egypt, Spain, France, French Guiana, French Lesser
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021168JUMP START, MODEL 970300
Medical Research Laboratories, Inc., Jun 26, 2002