Recall Z-1503-2009 ยท LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days

Arrow International Inc

Arrow Peripherally Inserted Central Catheter

DateJun 24, 2009 (initiated Feb 11, 2009)
StatusTerminated
ReasonLeakage: Material not lipid resistant, may crack and/or leak. Use of this device with lipid containing solutions or 70% isopropyl alcohol may cause leakage or air embolism.
Root causeDevice Design
ActionArrow International issued a letter dated 2/18/09 to all customers informing them of the problem and if a lipid resistant stopcock is required to contact the firm for replacements. New labels state: Warning - do not use with lipid containing solutions or 70% isopropyl alcohol, as these solutions may cause leakage or air embolism.
Quantity900,000 total, all products
DistributionWorldwide distribution: USA, Argentina, Australia, Bahamas, Belarus, Belgium, Brazil, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Lat