ClearanceBrief
Recall Z-1433-2009 ยท FPA Set, Administration, Intravascular

B. Braun Medical, Inc.

Ultrasite ADDitIV Primary IV Set with Universal Spike, Backcheck Valve, two Ultrasite Injection Sites and Spin-Lock Connector, Catalog # US3485. For use with Secondary Container.

DateJun 5, 2009 (initiated Mar 27, 2009)
StatusTerminated
ReasonDevice defect is occlusion/no flow.
Root causeMixed-up of materials/components
ActionAn Urgent - Medical Device Recall notice was issued via US Postal Service Certified Mail with registered return receipt mail or FedEx to all customer/distributors in receipt of affected product on 3/27/09. Direct customers received letters addressed to Materials Management that informed them of this recall and provided instructions for disposition of affected product in their inventory. Distributors received the direct customer recall letter along with a cover letter outlining their responsibilities to execute the recall and notify their customers as requested. Distributors are to immediate
Quantity52,050 units
DistributionWorldwide Distribution -- USA and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K790062IV ADMINISTRATION SET WITH BACKCHECK
National Patent Development Corp., Apr 10, 1979
K955585V2 INJECTION SITE
B.Braun Medical, Inc., May 15, 1996