Recall Z-1155-2009 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Ge Healthcare

GE Healthcare Fluorostar 7900 Mobile Fluoroscopy System

DateSep 24, 2009 (initiated Feb 19, 2009)
StatusTerminated
ReasonThe alarm timer is reset after each exposure and therefore never reaches the five minute alarm state unless one exposure exceeds 5 minutes in duration.
Root causeRadiation Control for Health and Safety Act
ActionGE Healthcare sent a customer notification out to all sites that have potentially affected systems. GEHC Surgery will also install a software correction into the affected systems in order to resolve this issue at no charge to the customer. Until the GEHC OEC Fluorostar 7900 has received the corrective software installed, they are encouraging facilities to exercise caution in monitoring the amount of cumulative fluoroscopy time when the affected system is in use.
Quantity554 units
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043076GE OEC FLUOROSTAR
Ge Oec Medical Systems, Inc., Dec 22, 2004