Recall Z-1757-2010 ยท IOL Treadmill, Powered

Cardiac Science Corporation

Treadmill models TM55, TM65, and CR60 (brand names include Quinton, Burdick, and Cardiac Science).

DateJun 4, 2010 (initiated Mar 30, 2009)
StatusTerminated
Reason28 stress and rehab treadmills manufactured between September 2006 and January 2009 may not have received adequate electrical safety testing prior to being shipped from the factory. If these components break down in the field, a patient or operator could be exposed to excess touch currents on the surface of the product. Because safety testing may not have allowed for complete evaluation of the el
Root causeOther
ActionOn April 3, 2009 the recalling firms notified consignees by certified letter to domestic customers and via FedEx to international customers with delivery confirmation signature required. Customers were advised to remove treadmill from service until it is tested by a Field Service Engineer. For customers with questions or those that have not already scheduled a service call with Cardiac Science Field Service Engineers, please contact the firm at 1.800.426.0337, option 1.
Quantity28 treadmills
DistributionDistributed domestically (including Puerto Rico) to 20 end-user accounts and 2 distributor accounts; Distributed internationally to: Canada, Mexico, Portugal, and Saudi Arabia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003576Q-TEL RMS (REHABILITATION MANAGEMENT SYSTEM), MODEL 000503
Quinton, Inc., Jan 25, 2001
K082173Q-STRESS AND HEARTSTRIDE, MODELS QSTRESS/Q40 AND H4/H4S
Cardiac Science Corporation, Aug 26, 2008
K864013MODEL Q55XT POWERED TREADMILL
Quinton, Inc., Nov 14, 1986
K990866QUINTON MEDTRACK CR PLUS TREADMILL
Quinton, Inc., Apr 29, 1999