Recall Z-2011-2009 ยท ECX Cylinder, Compressed Gas, And Valve

Intermed Gas Products Corporation

Post Medical Valves VO870A-Y and VO870-A-Y-T PSI 2015, PSI 2015, CGA 870, Inlet .750 STDS, Gas services (Oxygen), PSI 2015, Aluminum Cylinders, InterMed Gas Products LLC Industrial Division, 2056 Vista Parkway West Palm Beach, FL 33411.

DateSep 11, 2009 (initiated Mar 27, 2009)
StatusTerminated
ReasonFirm received reports from customers that valves malfunctioned when installation and/or filling process. Customers forwarded pictures of the defect where the top portion of the packing nut broke off from the threaded portion of the net; which may cause the valve to fail and result in uncontrolled release of oxygen when the valve is in the open position.
Root causeComponent design/selection
ActionIntermed Gas Products Corporation notified customers of the affected product by phone beginning March 27, 2009 and followed up with a formal "Product Recall Notice" letter dated March 31, 2009. All affected product should be located and returned to the firm for replacement. For further information, contact Intermed Gas Products at 1-561-483-1364.
Quantity8,767
DistributionNationwide Distribution -- FL, SC, TX, MO, TN, NC, NY, GA, MN, AZ.