ClearanceBrief
Recall Z-2116-2009 ยท HSB Rod, Fixation, Intramedullary And Accessories

EBI, L.P.

Uniflex Ti Buttress Thread Screw; TI Screw Buttress Thread Trauma Fixation Systems, Rx only Biomet, 100 Interpace Parkway, Parsippany, NJ 07054 Intended to aid in the alignment and stabilization of fractures to the skeletal system until healing has occurred.

DateSep 21, 2009 (initiated Mar 27, 2009)
StatusTerminated
ReasonCertain lots of raw material contain inclusions that could potentially reduce the strength of some devices.
Root causeNonconforming Material/Component
ActionLetters to Biomet/EBI reps were sent on March 29, 2009 by US mail to alert them of the product recall and the affected parts that may be in their possession. They were also asked to notify each customer and locate all affected inventory. All letters advised of the recall and the specific lot numbers involved. A letter was sent to the hospital and physician that received affected parts and asked that implanting physicians make any medical decisions.
Quantity5
DistributionWorldwide Distribution -- USA and Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982953TITANIUM INTRAMEDULLARY NAILS-VARIOUS STYLES
Biomet, Inc., Oct 21, 1998