GE Medical Systems, LLC

GE Healthcare, Drapes and Sterile Covers PART NUMBER DESCRIPTION 00-900352-01 DRAIN BAG W/HOSE (2000/2500) 2.0 PE MATERIAL 00-900352-01-OEC STRL, BAG, DRAIN WITH PLASTIC CLIPS AND HOSE 2000,2500 00-900352-02 STRL, BAG, DRAIN WITH METAL TABS 00-900493-03 COVER DISPOSABLE X-RAY TUBE 00-900605-01 UROFAST DRAIN BAG (UROFAST ONLY) 00-900605

DateAug 14, 2009 (initiated Aug 29, 2008)
StatusTerminated
ReasonGE Healthcare has found that the sterile products manufactured by Contour Fabricators, Inc. (CFI) for GE healthcare (GEHC) may have sterile seals that could be compromised during the shelf life of the products. This issue could result in the product not maintaining its sterile state.
Root causeRelease of Material/Component prior to receiving test results
ActionConsignees were sent 8/29/08 a GE Healthcare "Urgent Recall Notice" dated August 15, 2008. The letter was addressed to Director / Manager of Radiology and Director of Surgery. The letter described the problem and provided a list of products, solutions and requested the return of the attached fax forms. In the event of any questions or concerns, a service team should be contacted at 800-874-7378 option #8.
Quantity137,689 (123,373 US; 14,316 OUS)
DistributionWorldwide Distribution -- USA, Canada, Australia, Bahamas, China, Hong Kong, Italy, Japan, Korea, Mexico, Malaysia, Singapore, and El Salvador.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021049OEC 9800 PLUS DIGITAL MOBILE IMAGING SYSTEM
Ge Oec Medical Systems, Apr 17, 2002
K782155GENERAL EQUIPMENT COVER
Contour Fabricators, Inc., Jan 17, 1979
K820692BANDBAG, WRAP, STERILE
Contour Fabricators, Inc., May 3, 1982
K970185STERILE DISPOSABLE EQUIPMENT MAGNET COVER
Contour Fabricators, Inc., Apr 10, 1997
K993687UROVIEW 2800