Recall Z-1444-2009 ยท GEX Powered Laser Surgical Instrument

Sybaritic, Inc

Sybaritic, NannoLight MP50. Model Numbers and Description: 1.) 304207-MP 220 Volt, Multi-Platform configuration (able to attach optional laser handpiece when purchased separately), Subassembly (unit only, no safety glasses, owner's manual, etc. included); 2.) 404207 Completed unit (safety glasses, owner's manual, etc. incl

DateSep 3, 2009 (initiated Apr 12, 2008)
StatusTerminated
ReasonSybaritic Inc. had an issue with the 585nm filter used with the NannoLight MP50. The filter may cause unwanted burning and should not be used for vascular lesion treatments.
Root causeDevice Design
ActionA "Sybaritic" notification letter dated April 15, 2008 was issued to consignees. The letter described the modifications to the protocols for the NannoLight MP50 and the temporary replacement of 585 Filter for the 560 Filter for Vascular Lesions until further notice. They requested the immediate use of the protocols replacing all other versions. A follow-up letter dated June 3, 2008 was issued to customers. The letter included an attachment " Product Notice Nannolight MP50 585 nm "special" filter from Sybaritic, Inc. The letter described the new 585 nm "bandwidth filter". Also the notific
Quantity59
DistributionWorldwide Distribution -- US (states of WA, CA, MN, FL and AR), POLAND, THAILAND, CHILE, CANADA, KOREA, AZERBAIJAN, INDIA, SPAIN, ARGENTINA, RUSSIA, GREECE, PHILIPPINES, POLAND, INDIA, JORDAN, TURKEY, MEXICO, URUGUAY,BELGIUM, LEBANON, BULGARIA and SINGAPORE.