Recall Z-1325-2009 ยท NAW Microspheres Radionuclide

MDS Canada Inc. DBA MDS Nordion

TheraSphere Administration Set (Part Number K125914-034). TheraSphere is indicated for radiation treatment or as a neoadjuvant to surgery or transplantation in patients with unresectable Hepatocellular carcinoma(HCC) who can have placement of appropriately positioned hepatic arterial catheters. The device is also indicated for HCC patients with partial or branch portal vein thrombosis/occlusion.

DateJun 5, 2009 (initiated Feb 24, 2009)
StatusTerminated
ReasonInstructions for the use of the pinch clamp that is an integral part of the TheraSphere Administration Set were not included with the distribution of the device.
Root causeError in labeling
ActionA User Bulletin (TS-2009-1) dated February 23, 2009 was issued to all TheraSphere user sites. The bulletin described the issue, affected units, and recommended actions for users. The bulletin instructed users on the proper use of the pinch clamp. MDS Nordion also updated the package insert with these instructions. For questions about the recall, customers in North America contact MDS Nordion Customer Service at 1-866-363-3330 and customers outside North America call 1-877-474-5999 and request Customer Service. Customers may also e-mail MDS Nordion at TheraSphere@mdsinc.com.
Quantity26
DistributionWorldwide Distribution -- US (states of AZ, CA, CO, FL, IL, IN, KS, MD, MI, NC, RI, OH, OR, PA, WI) and country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.