Recall Z-1646-2009 ยท LZG Pump, Infusion, Insulin

Disetronic Medical Systems, Inc.

Roche ACCU-CHEK Spirit insulin pump. Part numbers 4540492001, 4759702001 and 4759729001.

DateMay 26, 2009 (initiated Apr 30, 2009)
StatusTerminated
ReasonThe "up" and "down" buttons, which are used for changing the program in the menu or to administer additional insulin through a bolus delivery, may experience intermittent failure or completely cease to operate.
Root causeDevice Design
ActionThe firm issued a press release on 4/30/09 and sent letters with attached Urgent Medical Device Recall (Field Corrective Action) notification dated 5/4/09 to distributors, health care providers and users. Disetronic will replace pumps within the affected serial numbers (SN02119552 to SN10006093) that have intermittent or complete loss of function of the up and/or down button promptly upon contact from the customer with a new pump that is not affected by the potential button failure. For pumps within the affected serial numbers (SN02119552 to SN10006093) which are functioning properly, no repl
Quantity14,929
DistributionWorldwide Distribution -- USA and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060876ACCU-CHEK INSULIN PUMP CONFIGURATION SOFTWARE (STANDARD), MODEL 04625137001; ACCU-CHEK INSULIN PUMP CONFIGURATION SOFTWA
Disetronic Medical Systems AG, Jun 15, 2006