ClearanceBrief
Recall Z-1727-2009 ยท KGI Densitometer, Bone

Hologic, Inc.

Hologic QDR 4500 series Bone Densitometer Software versions 12.6, 2.3., or 2.4 Indicated for the estimation of bone mineral density (BMD), comparison of measurable variables obtained from a given QDR scan to a database of reference values, the estimation of fracture risk, vertebral deformity assessment, body composition analysis, and discrimination of bone from prosthetics.

DateAug 25, 2009 (initiated Apr 7, 2009)
StatusTerminated
ReasonSystem may not turn off at the end of the x-ray and cause additional x-ray exposure
Root causeComponent design/selection
ActionHologic issued an Urgent: Medical Device Recall customer notification letter, dated April 17, 2009, by certified mail to inform customers of the problem. With the notification letter, customers will receive the following: -Software version Apex 12.6.2, 2.3.2, or 2.4.2 -Instructions to upgrade their systems with the provided software and to verify the new version on the system as a confirmation that the software was successfully loaded. -New Operator's Manual for installations where the operating characteristics of the system have changed due to the software upgrade. For Customers with systems
Quantity1914 units total
DistributionWorldwide Distribution -- USA, Canada, Europe, Austria, Belgium, Bulgaria, Czech Republic, Denmark, France, Germany, Great Britain, Greece, Hungary, Iceland, Italy, Latvia, Lithuania, Macedonia, Netherlands, Norway, Poland, Portugal, Romania, Russia, Serbia, Slovak Republic, Slovenia, Spain, Sweden,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072847APEX 2.0 SOFTWARE FOR QDR X-RAY BONE DENSITOMETERS
Hologic, Inc., Mar 28, 2008