ClearanceBrief
Recall Z-1343-2009 ยท FGE Stents, Drains And Dilators For The Biliary Ducts

Boston Scientific Corporation

Boston Scientific, Flexima and Percuflex Drainage Catheter Family (Pigtail Style) UPN# DESCRIPTION M001231360 IE TEMP TIP 8/22 M001231370 IE TEMP TIP 8/24 M001231380 IE TEMP TIP 8/26 M001231390 IE TEMP TIP 8/28 M001231400 IE TEMP TIP 10/22 M001231410 IE TEMP TIP 10/24 M001231420 IE TEMP TIP 10/26 M001231430 IE TEMP

DateMay 21, 2009 (initiated Mar 18, 2009)
StatusTerminated
ReasonBoston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the patient.
Root causeLabeling False and Misleading
ActionConsignees were sent a Boston Scientific " Urgent Medical Device Field Correction Notification" letter dated March 18, 2009. The letter was addressed to "The Hospital Field Action Contact". The letter described the problem, product affected and the revised catheter removal instructions. They also requested the return of the Acknowledgement Form.
Quantity586,276 US; 276,013 OUS total for all products
DistributionWorldwide Distribution -- USA including DC, Puerto Rico and Guam and countries of United Arab Emirates Austria, Bosnia and Herzegovina, Bahrain, Switzerland, Germany, Egypt, Spain, Finland, France, Great Britain, Ireland, Italy, Kuwait, Netherland, Pakistan, Portugal, Saudi Arabia, Sweden, Slovak Re
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K900205DRAINAGE CATHETERS W/A DISSOLVABLE TIP
Boston Scientific Corp, Aug 20, 1991