Recall Z-1393-2009 ยท JQP Calculator/Data Processing Module, For Clinical Use
Roche Diagnostics Corp.
Roche cobas 6000 analyzer series (including cobas c 501 and cobas e 601 modules) Middleware Solutions Instrument Manager Standard as follows: v.8.04.02.10, catalog no. 04785550001; v.8.04.03, catalog no. 04785550002; and v.8.06.02, catalog no. 05108993001. The device is used to interface hospital LIS systems with Roche c 6000 analyzers and allows order and result communication to occur between th
| Date | May 20, 2009 (initiated Apr 1, 2009) |
| Status | Terminated |
| Reason | Some data alarm flags do not transfer from the analyzer to the laboratory information system, since not all data flags are recognized by the instrument manager/middleware. |
| Root cause | Component design/selection |
| Action | Consignees were notified by "Urgent Medical Device Correction" letter dated April 1, 2009 and instructed to add error codes to the Instrument Manager driver and to contact the manufacturer of their LIS for instructions on adding data alarm codes to the LIS. Consignees were also instructed to complete the attached fax form after adding the error codes to their system and fax to 1-888-345-4690. Contact Roche Diagnostics Technical Support at 1-800-428-2336 for questions about information contained in the notification letter. |
| Quantity | 498 of all products |
| Distribution | Nationwide Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.