Recall Z-1389-2009 ยท MKJ Automated External Defibrillators (Non-Wearable)

Welch Allyn Protocol, Inc

Welch Allyn AED 10 automated external defibrillator and MRL Jumpstart (collectively "AED 10")

DateJun 23, 2009 (initiated Feb 26, 2009)
StatusTerminated
ReasonIssue #1: AED 10 displayed a "Do Not Use" symbol as a result of a blown F1 fuse. Issue #2: AED 10 models that incorporate a spring connector with the speaker experienced intermittent electrical connection to the circuit board causing a loss of voice prompts.
Root causeComponent design/selection
ActionWelch Allyn sent an Urgent Medical Device Recall letter dated February 26, 2009 to all distributors and customers, describing 5 failure modes. Customers were instructed to respond to the letter by calling 888-345-5356, and decide on replacement or discounted new purchase.
QuantityIssue #1: 4568 units; Issue#2: 5186 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021168JUMP START, MODEL 970300
Medical Research Laboratories, Inc., Jun 26, 2002