ClearanceBrief
Recall Z-1243-2009 ยท DYG Catheter, Flow Directed

Edwards Lifesciences, Llc

Edwards Lifesciences Swan-Ganz CCOombo CCO/SvO2/VIP Thermodilution Catheter with AMC THROMBOSHIELD (An Antimicrobial * Heparin Coating) For use with Vigilance Monitor. This package contains: One catheter and one volume-limited syringe. REF: 746HF8 For use inpatients who require hemodynamic monitoring. Intended to be used in combination with clinical pressure monitoring equipment to measure rig

DateApr 17, 2009 (initiated Feb 9, 2009)
StatusTerminated
ReasonCardiac output measurements were unavailable and the following message is displayed in the Vigilance monitor: "Fault. CCO Catheter Verification. Use Bolus Mode".
Root causeSoftware design
ActionEdwards initiated the worldwide recall via Urgent Customer Notification letters dated February 9, 2009 and are being sent by Federal Express starting on February 10, 2009. These letters are requesting that customers cease using Swan-Ganz" Continuous Cardiac Output (CCO) Thermodilution Catheter products from the indicated lots and that they return all unused product. Edwards Lifesciences requested that the customer provide written verification of stock and identify any unused product to be returned. Questions should be directed to Technical Service at 800-822-9837.
Quantity5,756 units
DistributionWorldwide Distribution -- AL, AR, AZ, CA, CT, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Hawaii, and Puerto Rico and countries of Canada, Austria, Belgium, Switzerland, Germany, Denmark, Spain, Finland, France,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K934742SWAN-GANZ CONTINUOUS CARDIAC OUTPUT/OXIMETRY/VENOUS PORT
Baxter Healthcare Corp, Jan 24, 1994