Recall Z-1318-2009 ยท JAQ System, Applicator, Radionuclide, Remote-Controlled

Siemens Medical Solutions USA, Inc

Mammo Test Model number 10144185, x-ray guided stereotactic biopsy system

DateJun 1, 2009 (initiated Jan 21, 2009)
StatusTerminated
ReasonTable may unintentionally lift during procedure
Root causeNonconforming Material/Component
ActionSiemens Medical Solutions issued a Safety Advisory Notice dated 1/22/09 which informs the customers of the potential issues and recommends that they stop using their MammoTest System until the prone table has been repaired. Linear Medical visited the customers to repair the weld joint on the prone table by January 23, 2009.
Quantity4 units
DistributionUSA distribution to IL, NY, SC, and WA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042095MAMMOTEST
Fischer Imaging Corp., Oct 29, 2004