Recall Z-1723-2009 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Medical Systems, LLC

Signa Openspeed 0.7T MR Systems, Model Numbers: Openspeed 3 (2138300-30), Openspeed 4 (2377062-5), Openspeed 5 (2377062-5, 2377062-30)

DateAug 25, 2009 (initiated Nov 3, 2008)
StatusTerminated
ReasonGE Healthcare has identified a potential pinch point hazard on Ovation MR Scanner when using CTL Array XL Body Flex and Opened Body coil. A patient's hand may be pinched between the magnet bore ceiling cover and the CTL Array, XL Body Flex, or Open Body coil if the patient puts his/her hand on the top of the coil. Two injuries and four complaints filed. 2/20/09: Recall being expanded to incl
Root causeDevice Design
ActionGE Healthcare issued a "Product Safety Notification" letter dated October 23, 2008. The letter was addressed to Hospital Administrators, Mangers of Radiology/Cardiology & Radiologists/Cardiologists. The letter described the Safety Issue, Affected Product, Safety Instructions, Product Correction and Contact Information for the Signa Ovation 0.35T Systems. A second GE Healthcare "Urgent Medical Device Correction" letter dated March 5, 2009 was sent 3/9/09 to consignees which addressed the expanded recall to the Signa OpenSpeed0.7T MR Systems. For further questions, contact your local GE Heal
Quantity170(156 US / 14 OUS)
DistributionWorldwide Distribution-- US (including states of AL, ALZ, AR, CA, CO,CT, DE, FL, FA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MT, MO, NE,NH, NV, NJ, NY, NC, OH, OK, OR, PA, PR, SC SD, TN, TX, VA, WA, WV and WI), CHINA, CYPRUS, ARGENTINA, BRAZIL EGYPT, DENMARK, FEDERAL REPUBLIC OF GERMANY, FEDERAL
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032795GE 0.7T SIGNA OPENSPEED WITH EXCITE MR SYSTEM
Ge Medical Systems, LLC, Oct 8, 2003