Recall Z-1208-2009 ยท LXH Orthopedic Manual Surgical Instrument

Wright Medical Technology Inc

Screwdriver for 3.5mm Screws, DARCO system, instrument for use in orthopaedic implant surgery. REF: DC4261, Non-sterile, Rx only, 1 each, Wright Medical Technology, Inc., 5677 Airline, Road, Arlington, TN 38002

DateApr 28, 2009 (initiated Feb 19, 2009)
StatusTerminated
ReasonThe package insert lists 3 sterilization procedures, 2 of which are not validated. Pre-vacuum steam sterilization is validated.
Root causeProcess design
ActionWright Medical notified their distributors of the recall on 03/11/2009 by letter explaining the problem and requesting return of the product from their inventory and from their consignees. The letter included a "fax back" response page to ensure the receipt of the notice. The hospitals and surgeons were notified of the recall also on 03/11/2009. All letters were delivered by Fed Ex.
Quantity682 devices
DistributionWorldwide distribution: USA, Canada, Germany, France, Belgium, the Netherlands, Italy, Finland, Austria, Denmark, Norway, Turkey, South Africa, United Kingdom, and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.